For full functionality of this site it is necessary to enable JavaScript.

Powertek SD1000 phase meter

Powertekuk SD1000 phase meter
Manufacturer: Powertekuk
Model: SD1000

Details

Datasheet

  0.02º phase accuracy and 0.001º resolution
bullet Fully isolated inputs
bullet 700 kHz frequency range
bullet Accurate with noisy, distorted and non-sinusoidal signals
bullet Fully auto-ranging and easy to use
bullet Voltage or current inputs
bullet High common-mode noise rejection
bullet 500 Vpk input voltage capability
bullet 1mV sensitivity
bullet 10000:1 input ratio / imbalance capability
bullet IEEE-488, RS232 and printer interfaces
bullet Traceable to international standards
A phase meter not influenced by input noise and distortion
The SD1000 phase meter offers superb accuracy, over a wide range of signal conditions.   Conventional phase meters can become erratic when small levels of noise and distortion are present - the result is often unstable and incorrect phase readings. The SD1000 overcomes this by using Discrete Fourier Analysis (DFT) - to reject any noise and distortion without the need for tracking filters.
 

Works with multiple zero crossings due to switching noise

Works with noise & multiple zero crossings

Synthesized waveforms such as DAC outputs

Works with digitally generated signals


Isolated Inputs
The SD1000 phase meter has totally isolated (floating) measurement inputs, allowing the user to connect anywhere in the circuit. This is very useful when the circuit under test does not have common grounds, eliminating accidental grounding and ground-loop currents.  The user may choose voltage-voltage phase measurement, voltage to current phase measurement or current to current phase measurement, ideal for phase protection, power meters, energy meters and current transformer calibration

World-wide Traceability
All measurements made with the SD1000 phase meter are traceable to National and International standards through the measurement standards of Powertek. The SD1000 is supplied with a certificate of conformance necessary for quality assurance standards such as ISO9001. A UKAS (independent measurement certification organisation) certificate is available.


Phase meter applications

bullet

Calibration laboratories

bullet

Phase meter

bullet

Single phase angle indicators V-A

bullet

Three phase phase angle indicators V-A

bullet

Three phase angle indicators V-A

bullet

Cos Phi indicators

bullet

Power factor meters

bullet

Watt meters

bullet

Multifunction watt transducers

bullet

Phase protection relays

bullet

Power meters

bullet

Energy meters

bullet

Current transformer calibration

bullet

Current probe phase delay

bullet

Transformer phasing

bullet

Synchro

bullet

Resolver Angle indicators


Verifying phase angle error of current transformers and current sensors




Non-inductive shunt Option 02F/B,
25A - 50A Models used to enable SD1000 phase angle measurements
between voltage and current



Non-inductive shunt Option 02F/B 1A - 20A
Models used to enable SD1000 phase angle
measurements between voltage and current


SD1000 phase meter specifications

Phase Range ± 180.000º or 0.00º to +360.00º
Accuracy (23º ± 5ºC; 1-year cal interval)   0.5Hz to 100Hz ± 0.02º ± 1 digit
>100Hz to 1kHz ± 0.03º ± 0.005º /kHz ± 1 digit
>1kHz to 100kHz ± 0.04º ± 0.005º /kHz ± 1 digit
>100kHz to 700kHz ± 0.25º ± 0.005º /kHz ± 1 digit
Resolution 0.001º
Repeatability 0.005º ( ±0.003º via interfaces)
Frequency Range 0.5Hz to 700 kHz
Accuracy ±0.1% (23ºC ± 5ºC)
Inputs Voltage Range (operational)
Current ranges using Options 02F/B
All voltage and current input combinations are possible
1mV rms to 350V rms (500Vpk)
20mA - 2Arms range Opt. 02F/B 1R00 5Hz - 20kHz
1.0Arms to 20Arms range Opt. 02F/B 0R01 5Hz - 20kHz
Bandwidth 0.5Hz to 700kHz
Isolation (inputs to ground) 2000V
Impedance 2 Mohm and 10pF
Environment Temperature Operating 0ºC to + 50ºC
Storage -40ºC to +70ºC
Humidity (non-condensing) 10% to 90% RH
Dielectric Strength Inputs to case or AC line input 2kV AC 50/60Hz 1 minute
Power supply to case 2kV AC 50/60Hz 1 minute
Power Requirements Line input voltage 110V or 220V AC ± 10%
Line input frequency 50/60Hz
Line input fuse 315mAT
Line input conformance IEC348 Class I; EN61000-3-2; EN61010 Class I; IEC801 Parts 1 to 4; EN55011
Power consumption 25VA
Mechanical Weight 5kg (6kg in delivery packaging)
Dimensions (W x H x D) 215mm x 144mm x 390mm

 

  • Quality Engagement