For full functionality of this site it is necessary to enable JavaScript.

FLUKE FLUKE-725/APAC/EMEA Multifunction Process Calibrator (Measure and simulate current, voltage, temperature, pressure)

  • Manufacturer:FLUKE
    Model: FLUKE-725/APAC/EMEA
    P/N: 5131140
    Origin: China
    Guarantee: 12 Month
  • Contact
    • Request a Quotation
  • Contact us for the best price

    Myanmar: 09 266993007

    Email: info@emin.com.mm

Measurement Accuracy

Voltage DC

30.000 V

0.02%+ 2 counts (upper display)

30.000 V

0.02%+ 2 counts (lower display)

100.00 mV

0.02%+ 2 counts

-10.00 mV to 75.00 mV

0.025 % + 1 count (via TC connector)

Current DC

24.000 mA:

0.02%+ 2 counts

Resistance

0.0 to 400.0 Ω

0.1Ω (4-wire), 0.15Ω (2- and 3-wire)

401 to 1500 Ω

0.5Ω (4-wire), 1Ω (2- and 3-wire)

1500 to 3200 Ω

1Ω (4-wire), 1.5Ω (2- and 3-wire)

Frequency

2.0 to 1000.0 CPM

0.05 % + 1 count

1.0 to 1100.0 Hz

0.05 % + 1 count

1.00 to 10.00 kHz

0.05 % + 1 count

Sensitivity

1 V peak-to-peak-minimum

Pressure

Accuracy

from 0.025% of range using any of 8 intrinsic safe pressure modules. (for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gauge, vacuum, absolute, dual and high pressure.

Source Accuracy

Voltage DC

100.00 mV

0.02%+2 counts

10.000 V

0.02%+2 counts

-10.00 mV to 75.00 mV

0.025 % + 1 count (via TC connector)

Current DC

24.000 mA (Source)

0.02%+ 2 counts

24.000 mA (Simulate)

0.02%+ 2 counts

Resistance

15.0 to 400.0 Ω

0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)

401 to 1500 Ω

0.5 Ω(excitation current 0.05 to 0.8 mA)

1500 to 3200 Ω

1 Ω(excitation current 0.05 to 0.4 mA)

Frequency

2.0 to 1000.0 CPM

0.05%

1.0 to 1100.0 Hz

0.05%

1.00 to 10.00 kHz

0.25%

Waveform:

5 V p-p squarewave, -0.1 V offset

RTDs and Thermocouples

Measure Accuracy:

NI-120
PT-100 (385)
PT-100 (393)
PT-100 (JIS)
PT-200 (385)
PT-500 (385)
PT-1000 (385)
Resolution:
J:
K:
T:
E:
R:
S:
B:
L:
U:
N:
Resolution
XK
BP

0.2 °C
0.33 °C
0.3 °C
0.3 °C
0.2 °C
0.3 °C
0.2 °C
0.1 °C
0.7 °C
0.8 °C
0.8 °C
0.7 °C
1.8 °C
1.5 °C
1.4 °C
0.7 °C
0.75 °C
0.9 °C
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F
0.6°C
1.2°C

Source Accuracy:

NI-120
PT-100 (385)
PT-100 (393)
PT-100 (JIS)
PT-200 (385)
PT-500 (385)
PT-1000 (385)
Resolution:
Note:
J:
K:
T:
E:
R:
S:
B:
L:
U:
N:
Resolution
Resolution
BP

0.2 °C
0.33 °C
0.3 °C
0.3 °C
0.2 °C
0.3 °C
0.2 °C
0.1 °C
Accuracy stated for 4-wire measurement.
0.7 °C
0.8 °C
0.8 °C
0.7 °C
1.4 °C
1.5 °C
1.4 °C
0.7 °C
0.75 °C
0.9 °C
J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C
0.6°C
1.2°C

Specifications

Ramp Functions

Source functions:

Voltage, current, resistance, frequency, temperature

Ramps

Slow ramp, Fast ramp, 25% step-ramp

Loop Power Function

Voltage:

24 V

Accuracy:

10%

Maximum current:

22 mA, short circuit protected

Step Functions

Source functions:

Voltage, current, resistance, frequency, temperature

Steps

25% of range, 100% of range

Environmental Specifications

Operating Temperature

-10 °C to +55 °C

Storage Temperature

-20 °C to 71 °C

Operating Altitude

3000 m

Safety Specifications

Agency Approvals

EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992

Mechanical & General Specifications

Size

130 x 236 x 61 mm

Weight

0.65 kg

Batteries

4 AA alkaline batteries

Battery Replacement

Separate battery compartment, accessable without breaking calibration seal

Side port connections

Pressure module connector, also used for remote real-time programming

 

Details

DATASHEET



Specifications: Fluke 725 Multifunction Process Calibrator

Measurement Accuracy
Voltage DC30.000 V0.02% + 2 counts (upper display)
30.000 V0.02% + 2 counts (lower display)
100.00 mV0.02% + 2 counts
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA0.02% + 2 counts
Resistance0.0 to 400.0 Ω0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency2.0 to 1000.0 CPM0.05% + 1 count
1.0 to 1100.0 Hz0.05% + 1 count
1.00 to 10.00 kHz0.05% + 1 count
Sensitivity1 V peak-to-peak-minimum
PressureAccuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC100.00 mV0.02% + 2 counts
10.000 V0.02% + 2 counts
-10.00 mV to 75.00 mV0.025% + 1 count (via TC connector)
Current DC24.000 mA (source)0.02% + 2 counts
24.000 mA (simulate)0.02% + 2 counts
Resistance15.0 to 400.0 Ω0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω1 Ω (excitation current 0.05 to 0.4 mA)
Frequency2.0 to 1000.0 CPM0.05%
1.0 to 1100.0 Hz0.05%
1.00 to 10.00 kHz0.25%
Waveform5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure accuracyNI-1200.2°C
PT-100 (385)0.33°C
PT-100 (393)0.3°C
PT-100 (JIS)0.3°C
PT-200 (385)0.2°C
PT-500 (385)0.3°C
PT-1000 (385)0.2°C
Resolution0.1°C
J0.7°C
K0.8°C
T0.8°C
E0.7°C
R1.8°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
ResolutionJ, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
XK0.6°C
BP1.2°C
Source accuracyNI-1200.2°C
PT-100 (385)0.33°C
PT-100 (393)0.3°C
PT-100 (JIS)0.3°C
PT-200 (385)0.2°C
PT-500 (385)0.3°C
PT-1000 (385)0.2°C
Resolution0.1°C
Accuracy stated for 4-wire measurement.
J0.7°C
K0.8°C
T0.8°C
E0.7°C
R1.4°C
S1.5°C
B1.4°C
L0.7°C
U0.75°C
N0.9°C
ResolutionJ, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
XK0.6°C
BP1.2°C
Specifications
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
RampsSlow ramp, Fast ramp, 25% step-ramp
Loop power functionVoltage24 V
Accuracy10%
Maximum current22 mA, short circuit protected
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Steps25% of range, 100% of range
Environmental Specifications
Operating temperature-10°C to +55°C
Storage temperature-20°C to 71°C
Operating altitude3000 m
Safety Specifications
Agency approvalsEN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical and General Specifications
Size130 x 236 x 61 mm
Weight0.65 kg
Batteries4 AA alkaline batteries
Battery replacementSeparate battery compartment, accessable without breaking calibration seal
Side port connectionsPressure module connector, also used for remote real-time programming
  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information