- Detector:
+ Gamma: CsI(Tl)
+ Neutron:X - Sensitivity:
+ for 137Сs, no less 100 cps/(µSv/h)
+ for 241Am, no less 250 cps/(µSv/h)
- Energy range:
+ gamma 0.033 - 3.0 MeV
+ Neutron:X
- Dose Rate :
+ gamma 0.01 – 130 µSv/h (1 µR/h – 13 mR/h)
+ Neutron: X - Dose: X
- Accuracy: ± 30% (at 137Cs)
- Response time: 0.25 s
- Radionuclide identification:
+ Special nuclear materials (SNM): 233U, 235U, 237Np, 239Pu (function available only upon customer request)
+ Medical radionuclides: 18F, 67Ga, 51Сr, 75Se, 89Sr, 99Mo, 99mTc,
103Pd, 111In, 123I, 131I, 153Sm, 201Ti, 133Xe
+ Naturally occurring radioactive materials (NORM): 40K, 226Ra, 232Th and daughters, 238U and daughters
+ Industrial radionuclides: 57Co, 60Co, 133Ba, 137Cs, 192Ir, 226Ra, 241Am
- Standards compliance: ITRAP (IAEA) requirements, ANSI N42.32 (most relevant parts), ANSI N42.33 (most relevant parts), ANSI N42.42
- Alarm type: Visual, audible, vibration
- Data recording: Up to 100 spectra
- Environmental protection: IP65
- Drop test on concrete floor: 1.5 m
- Power supply: one AA standard or rechargeable battery
- Battery life time: Up to 300 hours
- Operating temperature: -20oС to 50oС (-4oF to 122oF)
- Size (without holster): 130х60х46 mm (5 1/9 x2 1/3 x1 8/9 )
- Weight:
+ (without holster): 250 g (8.8 oz)
+ (with holster):X
- Low battery warning: LCD
- Overload indication: Visual and audible
+ Gamma: Visual and audial
+ Neutron: X
- PC Communication: USB