AC withstand voltage (ACW): 5kVac/100mA
DC withstand voltage (DCW): 6kVdc/20mA
Insulation resistance (IR): 3kVdc/50GQ
Ground bond resistance: None
Multi-channel scanning: Optional 8-channel high-voltage scanning card
Interface: 5.5-inch LCD and RS232/PLC/USB interfaces
Dimensions (WxHxD, mm): 426×132×520
AC withstand voltage test
Rated output capacity: 500VA (5,000V/100mA), short circuit current greater than 200mA
Output voltage setting
Range/accuracy: Range: (100-5,000)V, resolution: 1V, accuracy: ±(1%×Set value+5 digits)
Output frequency setting:
Range/accuracy: 50Hz/60Hz, accuracy: ±0.1Hz
Alarm current setting
Upper limit range/accuracy:
Range: (0.00-100.00)mA, resolution: 0.01mA, accuracy: ±(1%×Reading value+5 digits
Lower limit range/accuracy:
Range: (0.000-9.999)mA, resolution: 0.001mA, accuracy: ±(1%×Reading value+5 digits)
Time setting
Test time range : Range: 0, (0.5-999.9)s, 0 represents infinite duration, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits)
Ramp-up time range: Range: 0, (0.1-999.9)s, 0 represents off, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits)
Ramp-down time range: Range: 0, (0.1-999.9)s, 0 represents off, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits
Initial voltage setting: (0%-50%)×Output voltage set value
Current compensation setting: 0.000-10.00mA, total current+compensation current <100.00mA, automatic and manual
Arc detection: 1-9 (9 is most sensitive), 0 indicates the arc function is turned off
DC withstand voltage test
Rated output capacity: 120VA (6,000VDC/20mA)
Output voltage setting
Range/accuracy: Range: (100-6,000)VDC, resolution: 1V, accuracy: ±(1%×Set value+5 digits)
Alarm current setting
Upper limit range/accuracy:
Range: (0.0-20,000)µA, resolution: 0.1µA/1uA, accuracy: ±(1%×Reading value+5 digits
Lower limit range/accuracy: Range: (0.0-999.9)µΑ, resolution: 0.1µA, accuracy: ±(1%×Reading value+5 digits)
Time setting
Test time range: Range: 0, (0.5-999.9)s, 0 represents infinite duration, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits)
Ramp-up time range: Range: 0, (0.4-999.9)s, 0 represents off, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits)
Ramp-down time range: Range: 0, (1.0-999.9)s, 0 represents off, resolution: 0.1s, accuracy: ±(0.1%×Set value+2 digits)
Initial voltage setting: (0%-50%)×Output voltage set value
Charging lower limit current: (0-350.0)uA, automatic and manual
Arc detection: 1-9 (9 is most sensitive), 0 indicates the arc function is turned off
Insulation resistance test
Output voltage setting
Range/accuracy: Range: (100-3,000)VDC, resolution: 1V,accuracy: ±(1%×Set value+5 digits)
Ground bond resistance test
Rated output: Test current up to 40A and open circuit voltage below 12V.
Output current setting
Range/accuracy: Range: (2.0A-40.0A)AC, resolution: 0.1A, accuracy: ±(1%×Set value+2 digits)
Output voltage setting
Range/accuracy: Range: (3.0-10.0)VAC, resolution: 0.1V, accuracy: ±(1%xSet value+2 digits), under open circuit conditions