Atomtex BDKG-32 Dosimetric Gamma Radiation Detection Unit (γ, 30 nSv/h – 500 mSv/h)

Detector: Scintillation tissue-equivalent plastic, Ø70х80 mm

Energy range: 40 keV – 10 MeV

Measurement range of ambient dose equivalent rate: 30 nSv/h – 500 mSv/h

Measurement range of air kerma rate: -

Limit of intrinsic relative measurement error: ±20%

Energy dependence relative to 662 keV ( 137Cs): 

±25% ((40 keV – 3 MeV)

±40% (3 – 10 MeV)

Typical sensitivity to gamma radiation, cps/(µSv·h^-1 )

241Am: 8300

60Co: 850

137Cs: 1660

Response time for dose rate change from 0.1 to 1 µSv/h (µGy/h): ≤2 s[accuracy error ≤±10%]

Protection class: IР64

Interface: RS232

Power supply: RS232 – 5-9 VDC

Overall dimensions, weight: Ø80х245 mm, 0.78 kg