Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Fitting: 1/4” VCR
Communications / Connector : Profibus / Analog (Input 0 - 5 V; Output 0 - 5 V); 9-Pin Female D conn. / 15-Pin Male D conn.
Performance:
Full Scale Flow Range (N2Eq.): 3 sccm to 55 slm
Flow Accuracy: ±1% S.P. 35 - 100%, ±0.35% F.S. 2 - 35%
Repeatability & Reproducibility: <±0.2% S.P.
Linearity: ±0.5% F.S. (included in accuracy)
Response Time (Settling Time): Normally Closed Valve <1 sec. (within 2% for steps 0 - 10 through 0 - 100%)
Control Range: 2 - 100%
Number of Bins: 11 bins
Valve Shut Down: <1% of F.S.
Zero Stability: <±0.5% F.S. per year
Pressure Coefficient: 0.03% per psi (0 - 50 psi N2)
Attitude Sensitivity: <0.25% span change @ 90°C after rezeroing (N2 @ 50 psi)
Auto shut-off: The Auto Shut-off feature closes the GF80 valve when the set point drops below 1.5% of full scale
Available Gases: MultiFlo™ Capable
Ratings
Operating Temperature Range: 5 - 50°C (41 - 122 °F)
Maximum Operating Pressure1 150 psig (10 bar)
Differential Pressure Range:
3 - 860 sccm = 7 - 45 psid,
861 - 7200 sccm = 15 - 45 psid,
7201 - 50000 sccm = 25 - 45 psid
Typical pressure drop, high density gases like Argon gas applications require an additional 10 psid differential pressure
Leak Integrity (external): 1x10^-10 atm. cc/sec He
For more details, see datasheet.