BSI GDRM Gamma Dose Rate Monitor (24.6 keV ~ 10MeV)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Radiation Type: X and Gamma - from 24.6 keV to 10MeV

Measured quantities: Ambient dose equivalent rate H*(10), Ambient dose equivalent H*(10). Not for pulsed radiation fields

Measuring range, dose rate: 0.05 mkSv/h ÷ 5 Sv/h (up to 100 Sv/h)

Measurement principle: Dead time independent

Integration time: 10 s … 6.55 · 105 s (configurable)

Overexposure: 50 Sv/h over 5 min without decrease in counting rate

Response time: ≤ 8 s (1 μSv/h … 10 mSv/h, peak finder activated)

: ≤ 2 s (10 mSv/h … 5 Sv/h, peak finder activated)

Data interfaces: RS 485, RS 232, option USB

Operating temperature: −30 °C to +60 °C

Power supply: 10 ÷ 30 VDC, 1.5W

Protection: IP67 (including connector and filling valve)

Enclosure: Ø 100 x 400 mm

Weight: 1.5 kg