Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Bandwidth(-3dB): 100MHz
Rise time: ≤3.5ns
Accuracy: ±2%
Range selection (Attenuation rate): 100X/1000X
Maximum differential voltage(DC + Peak AC): 100X (±700V); 1000X (±7000V)
Common mode voltage (Vp) (DC + Peak AC): ±7000V
Maximum input voltage-to-earth (Vrms): 1000V CATIII; 2300V CATI*
Input impedance:
Single-ended to ground: 20MΩ
Between inputs: 40MΩ
Input capacitance:
Single-ended to ground: <5pF
Between inputs: <2.5pF
CMRR:
DC: >80dB
100kHz: >60dB
1MHz: >50dB
Noise(Vrms): 100X (<220mV); 1000X (<1.2V)
Differential overvoltage detection level: 100X (≥700V); 1000X (≥7000V)
Propagation time:
Probe: 100X (8.9ns); 1000X (6.6ns)
BNC Line(1m): about 5ns
Bandwidth limit filters (5MHz): ≥-3dB@5MHz
Overload indicator (red light): Yes
Overload alarm: Yes(Can shut up manually)
Automatic save: Yes
Offset setting function: Yes( Set in test mode)
Terminate load: ≥100kΩ
Power supply: USB 5V/1A adapter
Safety standard: IEC/EN 61010-031:2015 + AMD1:2018
EMC standard: EN61326-1:2013 EN61000-3-2:2006+A1:2009+A2:2009 EN61000-3-3:2013