Indicator: 50 ~ 8000 mg/m3
Measuring Range: 50 ~ 8000 mg/m3, Number of Pump Strokes 1
Sampling Time: 1.5 minute per pump stroke
Colour Change: White → Brown (ring)
Reaction Principle: Stoddard Solvent + I2O5 + H2S2O7 → I2
Coefficient of Variation: 15% (for 50 to 1000 mg/m3), 10% (for 1000 to 8000 mg/m3)
Nr. of tests: 10
Pump: GV-100S/110S
Catalog
Manual
Performance
| Measuring Range |
50 to 8000 mg/m3 |
| Number of Pump Strokes |
1 |
| Correction Factor |
1 |
| Sampling Time |
1.5 minutes per pump stroke |
| Detecting Limit |
5 mg/m3 (n=1) |
| Colour Change |
White → Brown (ring) |
| Reaction Principle |
Stoddard Solvent + I2O5 + H2S2O7 → I2 |
| Coefficient of Variation |
15% (for 50 to 1000 mg/m3), 10% (for 1000 to 8000 mg/m3) |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Change colour by itself to |
| Carbon monoxide |
≧0.1% |
+ |
Pale brown (whole layer) |
| Acetylene, Hexane, Ethylene |
≧0.2% |
+ |
Pale brown (whole layer) |
| Alcohols, Ketones,Esters |
|
No |
Pale brown (whole layer) |
| Aromatic hydrocarbons |
|
+ |
Dark green |
| Calibration gas generation |
Vaper pressure method |
| TLV-TWA |
TLV-STEL |
Explosive range |
| 100ppm(2014) |
- |
0.8 % or higher |