
GEO Calibration 2015-TS Portable Humidity Calibrator (5°C-55°C, 5 % ~ 95 % RH)
Manufacturer: GEO Calibration Model: 2015-TS - Contact
CALIBRATION
Traceability 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
Temperature: 18 °C to 28 °C
Humidity: Up to 80 % RH
STORAGE CONDITIONS
Temperature: -20 °C to 50 °C
Humidity: 0 % to 95 % RH (non-condensing)
Altitude: 2000 m
CONTROLLED RANGE
Temperature: 5 °C to 55 °C
Humidity @ 18°C : 5 % to 95 % RH
Humidity @ 35°C 5 % to 80 % RH
Humidity @ 23°C : 5 % to 95 % RH
Humidity @ 40°C 5 % to 75 % RH
Humidity @ 30°C: 5 % to 85 % RH
Humidity @ 50°C 5 % to 60 % RH
CALIBRATION SYSTEM ACCURACY
Temperature: +/- 0.10 °C Or Better Typically +/- 0.05 °C
Humidity: +/- 1.00 % RH Or Better Specially tuned systems can be as good as ± 0.60 %
Controller type : PID Controller
UNIFORMITY
Temperature : 0.10 °C Or Better Typically +/- 0.05 °C
Humidity : 0.30 % RH @ 18 °C to 28 °C Or Better Typically +/- 0.25 %
STABILITY
Temperature : 0.05 °C
Humidity: 0.15 % RH @ 18 °C to 28 °C
RAMP / SOAK RATE OF CHANGE
Depends on ambient conditions
Temperature
High → Low 1.50 °C / minute (Typical)
Low → High 5.00 °C / minute (Typical)
Humidity
High → Low 5.00 % RH / minute (Typical)
Low → High 10.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition 30 Minutes Maximum
Cold 30 Minutes
RESOLUTION
Display
Temperature 0.01 °C
Humidity 0.01 % RH
Dew Point 0.01 °C (Calculated)
USB
Temperature 0.01 °C
Humidity 0.01 % RH
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment