Spectral responsivity: 300 nm - 700 nm
Input optics: 5mm diffuser, cos FOV, f2 < 5%
Irradiance Range: max current X1: 0.2 mA
@270 nm: 22 W/cm²
@365 nm: 7.6 W/cm²
@385 nm: 6 W/cm²
@405 nm: 5 W/cm²
@435 nm: 4.2 W/cm²
@460 nm: 3.8 W/cm²
@700 nm: 12.8 W/cm²
- f2 (directional response/cosine error): f2 < 5%
- Calibration: Calibration of the irradiance sensitivity in A/(W/m²). Spectral calibration of all channels for measurements in the wavelength range (300 - 700) nm.
- Mounting: side M6 thread hole
- Connector: coaxial cable 2m long with ITT (-4) connector
- Dimensions: diameter: 37mm, height: 17 mm
- Humidity: 0 – 70% rH; non-condensing
- Temperature range:
+ Storage: (-10 to +50)°C
+ Usage: (+10 to +40)°C