SALINITY ACCORDING TO PPT (G/L)
LOW LADDER
Scale: 0.00 to 10.00 ppt (g/L)
Resolution: 0.01 ppt (g/L)
Accuracy: ±0.20 ppt (g/L)
HIGH LADDER
Scale: 0.0 to 70.0 ppt (g/L)
Resolution: 0.1 ppt (g/L)
Accuracy:
±1.0 ppt (g/L) (scale from 0.0 to 40.0)
±2.0 ppt (g/L) (range 40.0 to 70.0)
AUTO (DEFAULT)
The scale:
0.00 to 9.99 ppt (g/L)
10.0 to 70.0 ppt (g/L)
Resolution: 0.01 / 0.1 ppt (g/L)
Accuracy:
±0.20 ppt (g/L) (scale from 0.00 to 9.99)
±1.0 ppt (g/L) (scale from 0.0 to 40.0)
±2.0 ppt (g/L) (range 40.0 to 70.0)
SALINITY BY PSU
LOW LADDER
Scale: 0.00 to 10.00 PSU
Resolution: 0.01 PSU
Accuracy: ±0.20 PSU
HIGH LADDER
Scale: 0.0 to 70.0 PSU
Resolution: 0.1 PSU
Accuracy
±1.0 PSU (scale from 0.0 to 40.0)
±2.0 PSU (range from 40.0 to 70.0)
AUTO (DEFAULT)
The scale:
0.00 to 9.99 PSU
10.0 to 70.0 PSUs
Resolution: 0.01 / 0.1 PSU
Accuracy
±0.20 PSU (scale from 0.00 to 9.99)
±1.0 PSU (scale from 0.0 to 40.0)
±2.0 PSU (range from 40.0 to 70.0)
SALINITY ACCORDING TO S.G.
LOW LADDER
Scale: 1.000 to 1.007 S.G.
Resolution: 0.001 S.G.
Accuracy: ±0.001 S.G.
HIGH LADDER
Scale: 1,000 to 1,041 S.G.
Resolution: 0.001 S.G.
Accuracy: ±0.001 S.G.
AUTO (DEFAULT)
Scale: 1,000 to 1,041 S.G.
Resolution: 0.001 S.G.
Accuracy: ±0.001 S.G.
TEMPERATURE
Range: 0.0 to 50.0 oC (32.0 to 122.0 oF)
Resolution: 0.1 oC / 0.1 oF
Accuracy: ±0.5 oC / ±0.1 oF
Method
ppt: International Oceanographic Tables, 1966
PSU: Standard Methods for the Examination of Water and
Wastewater, 2520 B, Electrical Conductivity Method
S.G.: Standard Methods for the Examination of Water and
Wastewater, 2520 C, Density Method
Calibration: Automatic, 1 or 2 points at 5.00 ppt or 35.00 ppt
Temperature compensation: Automatic from 5.0 to 50.0 oC (41.0 to 122.0 oF)
Battery: 1 CR2032 3V Lithium-ion battery
Can be used for about 100 hours continuously
Automatically turns off
Customization: After 8 minutes, 60 minutes
Turn off the auto-off feature
Environment: 0 to 50 °C (32 °C to 122 °F); RH max 100%
Supply includes
- HI98326 meter
- Battery (in the device)
- Calibration solution 5.00 ppt (2 packs) and 35.00 ppt (2 packs)
- Protective cap
- User manual
- Plastic container