For full functionality of this site it is necessary to enable JavaScript.
EMIN.COM.MM
0
Product image

HORIBA PC2000-S pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

6,623,548 Kyat
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

pH

Range: -2.000 to 20.000 pH

Resolution: 0.1 / 0.01 / 0.001 pH

Accuracy: ± 0.003 pH

pH Buffer Groups: USA, NIST, NIST2, DIN, Custom

Calibration Points: Up to 5 (USA, NIST, NIST2, Custom) / Up to 6 (DIN)

ORP

Range: ± 2000.0 mV

Resolution: 0.1 mV

Accuracy: ± 0.2 mV

Calibration Option: Yes (Up to ± 200 mV)

Ion

Range: 0.000 µg/L to 9999 g/L

Units: µg/L ↔ mg/L ↔ g/L, ppm ↔ ppt, mmol/L ↔ mol/L

Resolution: 4 Significant digits

Accuracy: ± 0.3% full scale or ± 0.2 mV, whichever is higher

Calibration Points: Up to 5

Conductivity

Range:

0.000 to 1.999 µS/cm (k = 0.1)

2.00 to 19.99 µS/cm (k = 0.1, 1)

20.0 to 199.9 µS/cm (k = 0.1, 1, 10)

200 to 1999 µS/cm (k = 0.1, 1, 10)

2.00 to 19.99 mS/cm (k = 0.1, 1, 10)

20.0 to 199.9 mS/cm (k = 1, 10)

0.200 to 2.000 S/cm (k = 10)

Units: Auto ranging S/cm, S/m (µS ↔ mS ↔ S)

Resolution: 0.05% full scale

Accuracy: ± 0.6% full scale; ± 1.5% full scale > 18.0 mS/cm

Reference Temperature: 15.0 to 30.0 °C (adjustable)

Temperature Coefficient: 0.00 to 10.00 % per °C (adjustable)

Cell Constants: 0.070 to 13.00 (adjustable)

Calibration Points: Up to 4 (Auto) / Up to 5 (Manual)

Resistivity

Range: 0.001 MΩ・cm to 20.0 MΩ・cm

Resolution: 0.5% full scale

Accuracy: ± 0.6% full scale; ± 1.5% full scale > 1.80 MΩ●cm

Total Dissolved Solids (TDS)

Range:

0.01 to 9.99 mg/L (ppm)

10.0 to 99.9 mg/L (ppm)

100 to 999 mg/L (ppm)

1.00 to 9.99 g/L (ppt)

10.0 to 100 g/L (ppt)

Resolution: 0.01, 0.1, 1 mg/L ↔ g/L (ppm ↔ ppt)

Accuracy: ± 0.1% full scale

TDS Curves: EN27888, 442, NaCl, Linear (0.40 to 1.00)

Salinity

Range: 0.0 to 100.0 ppt / 0.00 to 10.00 %

Resolution: 0.1 ppt / 0.01%

Accuracy:± 0.2% full scale

Salinity Curves:NaCl / Seawater

Calibration Option:Yes (Manual; 1 point)

Temperature

Range: -30.0 to 130.0 °C / -22.0 to 266.0 °F

Resolution: 0.1 °C / °F

Accuracy: ± 0.5 °C / ± 0.9 °F

Calibration Option: Yes (± 10.0 °C / ± 18.0 °F range in 0.1 °C increments)

Memory: 2000

Auto Data Log: Yes

Real-time Clock: Yes

Date & Time Stamp: Yes

Measurement Modes: Auto Stable / Auto Hold / Real Time

Offset & Slope Display: Yes (Segment & Average Slopes)

Calibration Alarm: Yes (Programmable: up to 90 days)

Auto Shut-Off : Yes (Programmable: up to 30 mins.)

Electrode Status: On screen display

Diagnostics: Yes

Password Setting: Yes

Software Upgrade: Yes

PC / Printer Communication: Phono jack (USB / RS232C)

Meter Inputs: 2 x BNC, 2 x phono (ATC), DC sockets

Display: 5" Custom LCD with backlight and 320 segments

Power Requirement: AC adaptor 100 - 240V, 50 - 60Hz

Dimensions & Weight: 155(L) x 150(W) x 67(H) mm, 770g

Includes:

PC2000

9615S-10D refillable, glass-body pH electrode with built-in temperature sensor

9382-10D Ti/Pt black plastic-body conductivity cell k=1.0 with built-in temperature sensor

502-S USA pH buffers kit

503-S Conductivity standard solutions kit

Brochure