Volume: 256L
Interior dimensions: Width 640 x Height 800 x Depth 500 mm
Exterior dimensions: Width 824 x Height 1183 x Depth 684 mm
Number of supply trays: 2
Maximum number of tray grooves: 9
Operating temperature range: Minimum 50°C above ambient temperature up to 800°C
Setpoint resolution: 0.10°C
Temperature sensors: Uses 2 Pt100 temperature probes according to DIN Class A standard, used for displaying operating temperature and alarm
Control panel (ControlCOCKPIT): Dual display. Multifunctional PID microprocessor controller with two high-resolution color TFT screens
Language settings on ControlCOCKPIT panel: German, English, French, Spanish, Polish, Czech, Hungarian
Adjustable countdown timer: From 1 minute to 99 days
HeatBALANCE function: Used to adjust the heating level to balance the top and bottom of the drying chamber within an adjustable range of -50% to 50%
SetpointWAIT function: Time is activated when the temperature reaches the set temperature
Calibration: Freely selectable for 3 temperature values
Adjustable parameters: Temperature (0°C or 0°F), air convection door position, time program, time zone
Sterilization: Fixed sterilization program (4 hours/160°C) for sterilizing the working chamber, no sterilization for samples
Natural air convection
Pre-heated air mixture adjustment: By adjusting the air intake cover by 10%
Program storage in case of power failure
Programs: AtmoCONTROL software on USB memory stick used for programming, managing, and transferring programs via Ethernet or USB interface
ControlCOCKPIT control panel: USB port for loading programs, reading values, activating user index setup functions
Data display: Display of saved data on the ControlCOCKPIT control panel (up to 10,000 values equivalent to about 1 week)
Ethernet interface: Located at the back of the device for reading recorded data, loading programs, and online recording
Temperature control: Mechanical temperature limit device TB, protection class 1 according to DIN 12 880, automatically shuts off when the temperature exceeds approximately 20°C above the set temperature, overheat control TWW, protection class 3.1, or adjustable temperature limit TWB, protection class 2, selectable on the screen.
Additional temperature control: Integrated over and under temperature control "AFS," automatically following the setpoint value within a predefined tolerance range; alarm in case of over and under temperature, heating function is cut off in case of over-temperature, cooling function automatically cuts off in case of under temperature
Microprocessor-controlled PID with self-diagnosis system
Alarm: Visual and audible
Door: Fully insulated stainless steel door with 2 locking points, inner glass door
Housing structure: Scratch-resistant, hard, and durable stainless steel, galvanized steel at the back.
Internal memory: Capable of storing data for at least 10 years
100% air safety compliance (100% AtmoSAFE)
Connections: Ethernet LAN, USB
Power supply: 230 V, 50/60 Hz
Power consumption: Approximately 1700 W
Ambient temperature: 5 – 40°C
Humidity: Maximum 80% rh, non-condensing
Included Accessories:
Calibration certificate at 370°C (provided by Memmert)
Stainless steel mesh trays, 2 pieces
USB with AtmoCONTROL software
User manual