Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Accuracy: ≤ ± [1 % +0.1 ppb]
IP Rating: IP68
Model Family: InPro 6900/6900i/6950i
CIP resistant: Yes
Response Time: T90 = < 90 s @ 25°C (77°F)(air to N2)
Sensor Length: 70 mm
Cable Connector: K8S (IP68)
Design Pressure: 174 psi (12 bar)
Detection limit: 0.1 ppb
O-Ring Material: Silicone
Operating Range: 0.1 ppb air sat.
Sensor Diameter: 12mm
Connector Design: Straight
Measurement Range: 0-100% O2 saturation
Process Connection: Pg 13.5
Measurement Parameter: Oxygen
Measuring Temp. Range: 32 °F – 176 °F (0 °C – 80 °C)
Measurement Technology: Amperometric/Polarographic
Mechanical Temp. Range: 23 °F – 249.8 °F (-5 °C – 121 °C)
ISM (Digital) or Analog: ISM (Digital)
Certificates & Approvals: Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx certificate, FDA/USP Class IV, EHEDG
Hazardous Area Approvals: ATEX/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb Ex ia IIIC T69°C/T81°C/T109°C/T161°C Da/Db FM: IS / I, II, III / 1 / ABCDEFG/ T6 Ta=60 °C - 53 800 002 Entity
Operating Pressure (Bar): 0.2–6 Bar
Operating Pressure (PSI): 0–87 PSI
Wetted Metallic Material: 316L stainless steel
Wetted membrane material: PTFE
Autoclavable/Sterilizable: Yes
Primary Media/Application: Food and Beverage
Sterilizable/Autoclavable: Yes
Measured Pressure Resistance: 0.2 bar – 9 bar
Hygienic/Biocompatibility Approvals: FDA/USP Class 7
Surface Roughnes of Wetted Material: N5/Ra16 (Ra=0.4μm/16μin)
Residual signal in oxygen free media: <0.025% of signal in ambient air
Sensor signal in air at 25 °C (77 °F): 2500 to 6000 nA