Accuracy: ≤ ± [1 % +0.1 ppb]
IP Rating: IP68
Model Family: InPro 6900/6900i/6950i
Response Time: T98 = < 90 s @ 25°C (77°F)(air to N2)
Sensor Length: 70 mm
Cable Connector: K8S (IP68)
Design Pressure: 174 psi (12 bar)
Detection limit: 1 ppb
Operating Range: 1 ppb - O2 sat.
Connector Design: Straight
Measurement Range: 0-100% O2 saturation
Process Connection: Pg 13.5
Measurement Parameter: Oxygen
Measuring Temp. Range: 32 °F – 176 °F (0 °C – 80 °C)
Measurement Technology: Amperometric/Polarographic
Mechanical Temp. Range: 23 °F – 249.8 °F (-5 °C – 121 °C)
ISM (Digital) or Analog: ISM (Digital)
Hazardous Area Approvals: ATEX/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb Ex ia IIIC T69°C/T81°C/T109°C/T161°C Da/Db FM: IS / I, II, III / 1 / ABCDEFG/ T6 Ta=60 °C - 53 800 002 Entity
Operating Pressure (Bar): 0.2–6 Bar
Operating Pressure (PSI): 0–87 PSI
Wetted Metallic Material: 316L stainless steel
Wetted membrane material: PTFE
Autoclavable/Sterilizable: Yes
Primary Media/Application: Food and Beverage
Sterilizable/Autoclavable: Yes
Measured Pressure Resistance: 0.2 bar – 9 bar
Hygienic/Biocompatibility Approvals: FDA/USP Class 7
Residual signal in oxygen free media: <0.025% of signal in ambient air
Sensor signal in air at 25 °C (77 °F): 250 to 500 nA