Earth/Ground Bond Tests
Test Voltage: Nominal 6V AC
Frequency: 50 or 60 HZ (Independent of supply)
Display Range and Accuracy: 0-1500mΩ ±2% ±5 Counts
Display Resolution: 1mΩ
Compliance Test Ranges:
Current/Time*
4A - 110 secs
25A - 60 secs
40A - 30 secs
Test Duration limited Pro-Rata between 4A and 40A
Settable Output Current Range: 0.1A-40.0A
Selectable Range of Pass/Fail Levels: 0-1500mΩ
Insulation Resistance Test
DC Output Voltage: 0.10kV-6.00kV
(10V/Step when manually adjusted)
(50V/Step in test sequence program)
Display Range: 0.01MΩ -500MΩ
Display Range/Accuracy:
25kΩ - 100MΩ ± 5% ± 2 Digits
101MΩ - 350MΩ ± 10% ± 2 Digits
>350MΩ Indication Only
Display Resolution: 0.01MΩ
Pass/Fail Level: 0.00MΩ-500.0MΩ
AC Dielectric Strength Test
Programmable Voltage Range: 0.10kV - 5.00kV
(10V/Step when manually adjusted)
(50V/Step in test sequence program)
Frequency Independent of Supply: 50 or 60HZ
Voltage Display Range and Accuracy: 0.10kV-5.00kV ± 1% ±5 Counts
Voltage Display Resolution: 0.01kV
Current Display Range and Accuracy: 0.01mA - 20.00mA ± 1% ±5 Counts
Current Display Resolution: 0.01mA
Selectable Range of Pass/Fail Levels: 0.01mA-20.00mA
Maximum Current Output: 20.00mA @5kV
Optional Arc Detection: 9 Levels
DC Dielectric Strength Test
Programmable Voltage Range: 0.10kV - 6.00kV
(10V/Step when manually adjusted)
(50V/Step in test sequence program)
Voltage Display Range and Accuracy: 0.10kV - 6.00kV ± 1% ±5 Counts
Voltage Display Resolution: 0.01kV
Current Display Range and Accuracy: 0.01mA - 10.00mA ± 1% ±5 Counts
Current Display Resolution: 0.01mA
Selectable Range of Pass/Fail Levels: 0.01mA - 10.00mA
Maximum Current Output: 10.00mA
Power Output Rating
Test Voltage: 110V - 230V AC, 20A nominal (VIn-Vout)
Maximum Power Output: Up to 5.0kVA (dependent on mains supply rating)
Output Power Measurement: Single Phase Power Measurement
Display Range and Accuracy:
000.0VA to 99.9VA ± 5% ±5 digits
100.0VA to 999.9VA ± 5% ±2 digits
Selectable Range of Pass/Fail Levels: 000.2 - 999.9VA
Single Phase Power Factor Measurement
Display Range: 0.000 - 1.000* *Displayed value for indication only
Selectable Range Pass/Fail Levels: 0.000 - 1.000
Leakage Current Measurment
Display Range and Accuracy: 0.10mA - 20.00mA ± 1% ±5 Counts
Display Resolution: 0.01mA
Selectable Range of Pass/Fail Levels: 0.01mA - 20mA
Touch Current Measurement
Display Range and Accuracy: 0.02mA - 5.00mA ± 1% ±5 Counts
Display Resolution: 0.01mA
Selectable Range of Pass/Fail Levels: 0.02mA - 5.00mA
Measuring Device: IEC/EN 60990 Fig 4
Case Dimensions and Weight
Case Dimensions: 200W x 300H x 370D mm
Weight: 15kg approx
Operating Voltage Range: Selectable 115V - 230V AC RMS, 50/60Hz
Standard Accessories:
Instruction Manual
Power Cord
Guard Plug
Comms Cable