Temperature range (*1): +15℃ to +55℃ (room temperature -10℃ to room temperature +30℃)
Shaking motion: Switchable Reciprocal/Orbital shaking
Temp. control accuracy: ±0.3℃
Speed range: 20 to 300 r/min (200 r/min at the upper of stacked units )
Shaking width: 25mm(≒1inch)
Capacity:
Universal shaking platform MT-4430: Erlenmeyer flask 500 mL × 8 pcs
Universal shaking platform MT-4030: Erlenmeyer flask 500 mL × 6 or 5 L × 1
Door Style: Single swing door (Leftward open, Can be changed to rightward open) (*2)
Maximum load: Approx. 7 kg (Platform included)
Ambient temp. range (*3): +5℃ to +30℃
Heating cooling method: Peltier element
Program functions: 1 program, 9 segments: Temp., Shaking speed and time/per segment (Setting range: 00 h 00 min to 99 h 59 min), Temp. gradient function (*4)
Safety devices/functions: Leakage/Overcurrent breaker, Fuse, Alarm of abnormal sample temp. (Set as desired), Detecting the door opening and closing, Self-diagnosis,function for sensor error, Motor current limiter
Dimensions (W × D × H): 600 × 732 × 643 mm
Weight: Approx. 86 kg
Power supply (*5): AC100V/AC220-240V /7A
Standard accessories: Power cable ×1pc
(*1)The value at 25℃ of the rear surface intake temp. The temperature range of the unit is 10℃ below RT to 30℃ above RT, or within the operating temperature range. When using the LC-450EXP LED irradiation unit, the temperature range is +20℃ to +50℃ (room temperature -5℃ to room temperature +25℃). If a low temperature of +4℃ or +55℃ or higher is required, please refer to the existing mediumsized BR group.
(*2)Distributor can make necessary changes.
(*3)No condensation.
(*4)The transfer time of temperature can be set up.
(*5)200 V or 240 V specifications are also available by special order.