Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

- Performance:
Measurement Range: 0 to ±20 kV DC or peak AC
Measurement Accuracy: Better than ±0.1% of full scale, referred to the voltage monitor
Speed of Response (10% to 90%): Less than 200 µs for 1 kV step. Less than 5 ms for 20 kV step change
Full Signal Bandwidth: DC to better than 25 Hz
Stability:
Drift with Time Less than 100 ppm/hour, noncumulative
Drift with Temperature Less than 100 ppm/˚C
- Voltage Monitor:
Output: A buffered output provides a low-voltage replica of the measured voltage
Ratio: 1/1000th of the measured voltage
Output Noise: Less than 20 mV rms*
Output Impedance: Less than 0.1 W
- Voltage Display:
Voltage Display: 4 ½ digit LED display
Range: 0 to V ±19.99 k
Resolution: 1 V
Zero Offset: ±2 counts, referred to the voltage monitor
Sampling Rate: 3 readings per second
- Mechanical:
Dimensions: 230 mm H x 441 mm W x 432 mm D (9.06” H x 17.36” W x 17” D)
Weight: 17 kg (37 lb)
Voltage Monitor Output Connector: BNC connector
Ground Receptacle: Green binding post
- Operating Conditions:
Temperature: 0°C to 40°C (32°F to 104°F)
Relative Humidity: To 90%, noncondensing
Altitude: To 2000 m (6561.68 ft.)
- Electrical:
AC Line Cord Receptacle: Standard 3-prong with integral fuse holder
Line Voltage: 180 to 250 V AC at 48 to 63 Hz