- Resistance Specification:
Nominal resistance ( Ω ): 10
Initial deviation [1] (±μΩ/Ω ) @ (23±1)°C: 30
Annual Stability ( ±μΩ/Ω ): 12
Temperature coefficient ( ppm/K ): 2.0
Power factor [2] ( ±ppm/Rated power ): 1
- Current Measurement:
Nominal input current ( A ): 0.1
Nominal Output current ( V ): 1
AC/DC difference[3] ( k=2,±ppm ):
53 Hz: 20; 1 kHz: 25; 10 kHz: 25; 100 kHz: 25
Maximum current( A ): 0.2
Instantaneous current (<5s): 1.5 A
- Phase Displacement ( k=2, ±μrad ):
53Hz: 5; 400Hz: 5; 1kHz: 10; 10 kHz 30; 50kHz: 150; 100kHz: 300
- General Specifications:
Power supply: ——
Temperature performance:
Operating temperature: 13 °C ~ 33 °C;
Calibration temperature: 18°C~28°C;
Storage temperature: 5 °C to 45 °C
Note: When stored in a harsh temperature environment, the resistance value
may change temporarily and be restored to the original resistance value within
the following 30 days.
Humidity performance:
Operating humidity: < 50% r. H
Storage humidity :(15%~80%) R·H, no condensation