Xrite 361T Densitometer (0.0 D ~> 6.0 D; 0 % ~100 %)

Short Term Repeatability - Density: 

- Ortho,Visual (2 mm and 3 mm aperture):

  0.0 D to 5.0 D: ± 0.01 D

  5.0 D to 5.5 D: ± 2 %

  5.5 D to 6.0 D: ± 3 % 

- Ortho,Visual (1 mm aperture):

  0.0 D to 4.5 D: ± 0.01 D

  4.5 D to 5.0 D: ± 1 % 

- UV (3 mm aperture):

  0.0 D to 3.5 D: ± 0.01 D

  3.5 D to 4.0 D: ± 1 %

Measurement Geometry: Per ANSI PH 2.19, ISO 5/2

Inter-Instrument Agreement: 0.0 D - 5.0 D: ± 0.02 D, ± 2 %

Linearity:

- Ortho,Visual (2mm and 3mm aperture):

  0.0 D to 5.0 D: ± 0.02 D

  5.0 D to 5.5 D: ± 1%

  5.5 D to 6.0 D: ± 2% 

- Ortho,Visual (1mm aperture):

  0.0 D to 4.5 D: ± 0.02 D

  4.5 D to 5.0 D: ± 1 % 

- UV (3mm aperture):

  0.0 D to 3.5 D: ± 0.02 D

  3.5 D to 4.0 D: ± 1 %

Aperture: 1mm, 2mm and 3mm (0.5mm optional)

Humidity: ≤ 76%

Operating Temperature Range: 10°C to 40°C/50°F to 104°F

Storage Temperature Range:-20° to 50°C/-4° to 122°F

Dimensions (length, width, height): 43.5 cm x 33.0 cm x 15.2 cm/17.1" x 13.1" x 6.0"

Weight: 8.63 kg/19 lbs.

Shipped Weight: 17.5 kg/7.95 lbs.

Communication Interface: RS-232 serial interface

Instrument Display: 8 character LCD

Voltage: 90 - 130 VAC 50 - 60 Hz ; 120 - 260 VAC 50 - 60 Hz

Warm Up Time: 2 minutes; 5 minutes (UV)

Wattage: 80 VA max.

Density Range: 0.0 D to > 6.0 D; 0 % to 100 %

Density Standards: X-Rite Visual, X-Rite Ortho, X-Rite UV

Industry Standards: ISO 5-2, ANSI PH2.19