Measuring volume X = 700 mm, Y = 1000 mm, Z = 600 mm
Specifications here below hold for
S1: active sensors VAST XT gold, VAST XTR gold
S2: passive sensor VAST XXT with RDS
temperature range 18 - 22 °C
maximum temperature gradients 1.0 K/h; 1.5 K/d; 1.0 K/m
Max. Permiss. length measurement Error (MPE) as per ISO 10360-2:2009
MPE_E0 (3D) = (1.4 + L/350) μm for S1
MPE_E0 (3D) = (1.5 + L/350) μm for S2
Max. Permiss. Error (MPE) of scanning probing as per ISO 10360-4:2000
within Maximum Permissible Time (MPT) for scanning test
MPE_THP = 2.0 μm in MPT_tau = 40 s for S1
MPE_THP = 2.7 μm in MPT_tau = 50 s for S2
Single stylus form error as per ISO 10360-5:2010
MPE_PFTU = 1.4 μm for S1
MPE_PFTU = 1.5 μm for S2
Form measuring tolerance as per ISO 12181:2011 (VDI/VDE 2617 part)