Number of detectors : 1
Detector type: InGaAs and integrating cavity
Detector size: 9 mm input aperture
Wavelength range (nm) : 850 to 1660
Power range (dBm) (typical) : 37 to –55
Uncertainty : ±(5 % + 3 nW) c, e
Uniformity (dB) : ±0.05
Polarization‑dependent responsivity (dB) : ±0.008 typical
Linearity : ±0.11 dB (35 dBm to 30 dBm); ±0.05 dB (30 dBm to 5 dBm); ±0.015 dB (5 dBm to –22 dBm)
Wavelength resolution (nm) : 0.01
Stabilization time (ms) (typical) : 0.3
Sampling rate (sample/s/channel): Up to 5208
Trigger input voltage (Vdc): 0-5 (TTL-type) h
Analog output voltage (Vdc) (typical): N/A
Fiber type (µm) : 5/125 to 62.5/125