1) AC Hipot test
Rated Output: 5000V/40mA
Output Voltage: 100~5000V
Output Frequency: 50Hz/60Hz
Alarm Current
+ Upper limit: 0.10~40.00mA
+ Lower limit: 0.000~9.999mA
Current compensation: 0.000~40.00mA
2) DC Hipot test
Output Voltage: 100~6000Vdc
Alarm Current:
+ Upper limit: 0~10000μA
+ Lower limit: 0.0~999.9μA
Current compensation: 0 ~200.0μA
Discharge Time: ≤200ms
3) Insulation Resistance test
Output Voltage: 100~2500Vdc
Alarm Current
+ Upper limit: 100-499V,1~1000MΩ, 1000~2000MΩ; 500-2500V,1~199MΩ, 200~999MΩ 1000~9999MΩ
+ Lower limit: 100-499V,1~1000MΩ, 1000~2000MΩ; 500-2500V,1~199MΩ, 200~999MΩ 1000~9999MΩ
4) Ground Bond test
Rated Output: Max current 32A,Max test resistance:600mΩ,max open circuit voltage:12V
Output Current: 2.0~32.0A
Output voltage: 3.0~6.4V
Output Current: Frequency: Sin wave, 50/60Hz
5) Leakage Current test
Upper/Lower limit range: 0.0V~300.0V AC
Max loading curren: Max 20A
Power supply: Single phase output, AN1640H need operate with optional isolated power supply
6) Power test
Power upper/lower limit: 0.00W~99.99W、100.0W~999.9W、1000W~6000W
Voltage upper/lower limit: 0.00V~99.99V,100.0V~300.0V,Resolution:0.01V、0.1V
Current upper/lower limit: 0.00mA~99.99mA、100.0mA~999.9mA、1.000A~9.999A、10.00A~20.00A
Product Detail
AN1640H(F)/AN1651H(F) safety analyzer is mainly used to provide the test solution for electric car, grid, photovoltaic, electrical appliances, information equipment, medical equipment industry as well as innovative solutions for the various electrical measurement.
Main Features
★High Accuracy
The accuracy of ISA series safety performance is 1%, built-in power parameters accuracy is 0.2%
★Fast Test
ISA series safety analyzer support Ground bond and insulation resistance parallel test, double station asynchronous serial test, double station intelligence parallel test, and realize the test speed up to 4 times of that of the conventional safety instrument.
Unique quick response, all communication command can complete the data back within 0.1 s. Individual safety test can be finished in 0.5 s.
★Intelligent management
first based on the industrial development of android platform, all touch human-computer interaction. Support input smart reminder, extended function of Internet of things, support RFID and bar code scanning interface
★Quality Control
Data storage, data analysis and report generation,server data interface and so on, meeting the requirements of modern line.
Spec.
AC Hipot test |
|||
Rated Output |
200VA(5000V/40mA,short circuit current is greater than 200mA; |
||
Output Voltage |
Range/accuracy |
100~5000V,±(1%of setting+2 counts) |
|
Output Frequency |
Range/accuracy |
50Hz/60Hz,±0.1% of setting |
|
Alarm Current |
Upper limit/ accuracy |
0.10~40.00mA,±(1%of setting+2 counts) |
|
Lower limit/ accuracy |
0.000~9.999mA,±(1%of setting+2 counts) |
||
Start Voltage |
(0%~50%)× output voltage setting |
||
Current compensation |
0.000~40.00mA,Auto/Manual |
||
ARC detection |
1~9 level,9level is the most senstive,0 is off |
||
DC Hipot test |
|||
Output Voltage |
Range/accuracy |
100~6000Vdc,±(1%of setting+2 counts) |
|
Alarm Current |
Upper limit/ accuracy |
0~10000μA,±(1%of setting+2 counts) |
|
Lower limit/ accuracy |
0.0~999.9μA,±(1%of setting+2 counts)<span style="font-size:9.0pt;line-hei |