1) AC Output
Current range/Reeolution/Accuracy: 5.0~30.0A ,0.1A, ±(3% of setting+5 counts)
Voltage: <12V
Frequency: 50Hz/60Hz ±100ppM
2) Parameter Setting
Resistance set rang/Resolution: 1~600mΩ(below 10A), 1~300 mΩ(10-25A) 1~200 Mω(above 25A) 1 mΩ/step
Voltage set rang/Resolution: 0.01~6.00V(below 10A),0.01~7.5V(10~25A) 0.01~600V(above 25A) 0.01V/step
Time set range/Resolution: 1.0~999.9s(LLL=continuous), 0.1s/step
3) Measuring Function
Measurement voltage: 0.01~3.00V, ±(3% of reading+5 counts)
Resistance range/Accruacy: 1~600 mΩ, ±(3% of reading+5 counts) 300~7.50V, ±(3% of reading)
Max rated load: 180VA
Current fluctuation: ≤1%
Interface function: RS232(RS485 optional),Remote control
Dimension(W×H×D): 320×110×400mm
Product Detail
Main features
* Latest international standard: Comply with CCC,IEC,EN,VDE,BS,UL,JIB standard requrements.
* Result can be shown as resistance or voltage to satisfy different industrial standard requirements
* Measurement current 5-30A,output frequency 50/60HZ
* Leakage current upper setting,intelligent judgment
* VFD display,simple operation
* Remote,RS232/485 interface
Spec.
|
Model |
AN9613X(F) |
|
|
AC Output |
Current range/Reeolution/Accuracy |
5.0~30.0A ,0.1A,±(3% of setting+5 counts) |
|
Voltage |
<12V |
|
|
Frequency |
50Hz/60Hz ±100ppM |
|
|
Parameter Setting |
Resistance set rang/Resolution |
1~600mΩ(below 10A), 1~300 mΩ(10-25A) 1~200 Mω(above 25A) 1 mΩ/step |
|
Voltage set rang/Resolution |
0.01~6.00V(below 10A),0.01~7.5V(10~25A) 0.01~600V(above 25A) 0.01V/step |
|
|
Time set range/Resolution |
1.0~999.9s(LLL=continuous), 0.1s/step |
|
|
Measuring Function |
Measurement voltage |
0.01~3.00V, ±(3% of reading+5 counts) |
|
Resistance range/Accruacy |
1~600 mΩ, ±(3% of reading+5 counts) 300~7.50V, ±(3% of reading) |
|
|
Max rated load |
180VA |
|
|
Current fluctuation |
≤1% |
|
|
Interface function |
RS232(RS485 optional),Remote control |
|
|
Dimension(W×H×D) |
320×110×400mm |
|