Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

1) AC Hipot test
Booster rated output capacity: 500VA(5000V/100mA)
Booster maximum short-circuit current: >200mA
Output waveform: Sine wave
Output voltage setting: 500~5000V ± 3%
Frequency setting: 50Hz/60Hz ±0.1%
Current upper-limit: 0.10~99.99mA
Current lower-limit: 0.00~10.00mA
Compensation current : 0.00~10.00mA
Time setting: 1~999s
2) Insulation Resistance test
Rated output capacity: 2VA(1000V/2mA)
Output voltage setting: DC 500Vor1000V
Resistance upper-limit range: 0.0~99.9 MΩ,100~2000MΩ
Resistance lower-limit range: 0.3~99.9 MΩ, 100~2000MΩ
Insulation resistance measurement: 0.0~99.9 MΩ, 100~2000 MΩ
3) Ground Bond Resistance test
Output current setting: 3~30A
Current frequency setting: Sine waveform 50Hz/60Hz
4) Leakage Current test
Current upper-limit range: 0.050~9.999mA
Current lower-limit: 0.000~5.000mA
Leakage current compensation setting: 0.000~1.000mA
Leakage current measurement: 0.050~9.999mA
Output voltage measurement: 60~280V
Dimension (WxHxD): 483x600x1355
Gross weight: 80 kg
Product Detail
Main features
* Comply with CCC,IEC EN,VED,BS,UL,JIS standard requiements.
* Latest international standard:500VA AC Hipot capacity(AC 5KV/100mA),short-circuit output current exceeds 200mA.Built-in optional leakage current test network MD card,meeting requirements of different industrial standards Maximum test current of Ground Bond resistance is up to 64A
*6 in 1:AC Hipot,Insulation Resistance,Ground Bond resistance,Leakage Current,Running Power,Low-Voltage Start.
Standby power consumption is optional for AN9651F(F)and AN9651S(F),minimum is low to 0.50w,accuracy is 0.2%
* Simple operation:LED backlit screen display.menu operation,Chiense/English langues,socket test fixture.
* Flexible configuration:Built in isolated transformer,AC frequency conversion power supply,frequency conversion power supply and industrial PC combiantion
* Satisfy lattest requirements:Follow the production line developing trends of comprehensive,fast,automation and information.Satisfy quality management,5s management and refid management requirements.
|
AC Hipot test |
|||
|
Booster rated output capacity |
500VA(5000V/100mA) |
||
|
Booster maximum short-circuit current |
>200mA |
||
|
Output waveform |
Sine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load) |
||
|
Output voltage setting |
Range |
500~5000V |
|
|
Resolution |
5V |
||
|
Accuracy |
±(3%×setting+10V) |
||
|
Frequency setting |
Range |
50Hz/60Hz |
|
|
Accuracy |
±0.1%×setting |
||
|
Alarming limits setting |
Current upper-limit |
0.10~99.99mA |
|
|
Current lower-limit |
0.00~10.00mA |
||
|
Compensation current setting |
Range |
0.00~10.00mA, automatic testing, compensation can be set to ON/OFF |
|
|
Time setting |
Range |
1~999s |
|
|
Ramp up time range |
1~100s |
||
|
Accuracy |
±(0.1%×setting+1count) |
||
|
Breakdown current alarm |
Range |
0.10~99.99mA |
|
|
Accuracy |
±(3%×reading+3counts) |
||
|
Insulation Resistance test |
|||
|
Rated output capacity |
2VA(1000V/2mA) |
||
|
Output voltage regulation |
<3%(From no load to full load) |
||
|
Output voltage ripple |
<1% |
||
|
Output voltage setting |
Range |
DC 500Vor1000V |
|
|
Accuracy |
±(3%×setting+3counts) |
||
|
Alarming limits setting |
Resistance upper-limit range |
0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjudgment for upper limit) |
|
|
Resistance lower-limit range |
0.3~99.9 MΩ, 100~2000MΩ |
||
|
Time setting |
Range |
1~999s |
|
|
Accuracy |
±(0.1%×setting+1count) |
||
|
Insulation resistance measurement |
Range |
0.0~99.9 MΩ, 100~2000 MΩ |
|
|
Accuracy |
≤200MΩ: ±(3%×reading+3counts) |
||
|
>200MΩ:±(8%×reading+8counts) |
|||
|
Ground Bond Resistance test |
|||
|
No load output voltage |
<12v span=""> |
||
|
Output current setting |
Range |
3~30A |
|
|
Accuracy |
±(3%×setting+0.2A) |
||
|
Current range instruction |
Optional: 3 ~ 64A test current |
||
|
Current frequency setting |
Range |
Sine waveform 50Hz/60Hz |
|
|
Accuracy |
±0.1%×setting |
||
|
Test time setting |
Range |
1~999s |
|
|
Accuracy |
±(0.1%×setting+1count) |
||
|
Result display mode |
Ground resistance or voltage |
||
|
Ground resistance display mode |
|||
|
Alarming limits setting |
Resistance upper-limit range |
3A≤output current≤10A:10~600 mΩ |
|
|
Resistance lower-limit range |
0~100 mΩ |
||
|
Compensation resistance setting |
Range |
0~200 mΩ, automatic testing, compensation can be set to ON/OFF |
|
|
Ground bond resistance measurement |
Range |
10~600 mΩ |
|
|
Accuracy |
±(3%×reading+ |
||