Illuminating/Observation System: d/8゜(Diffused illumination, 8゜viewing angle)
SCI / SCE measurement (including specular reflection and eliminating specular reflection measurement mode), including UV / exclusion UV measurement;
Conform to: CIE No.15, ISO7724-1, ASTM E1164,DIN5033
Integrating sphere Size: Φ 48mm
Light Source: Combined LED sources
Spectroscopic Mode: Concave Grating
Sensor: 256-pixel dual-array CMOS image sensor
Wavelength range: 400~700nm
Wavelength interval: 10nm
Reflectance range: 0~200%
Illuminating/Measuring Aperture: Single Aperture: MAV(Big Aperture):Φ8mm/Φ10mm
Measurement Mode: SCI & SCE
Color Space: CIE LAB XYZ; Yxy; LCh; CIE LUV,Hunter LAB
Color difference Formula: AE*ab, ΔΕ*υν, ΔΕ 94, ΔΕ cmc (2:1),ДЕ*cmc (1:1),
ΔE*00,ΔE
Other Chromaticity Data: WI(ASTM E313, CIE/ISO, AATCC, Hunter), YI(ASTM D1925, ASTM 313), TI(ASTM E313, CIE/ISO), Metamerism Index (Mt), Color Stain, Color Fastness
Observer: 2°/10°
Illuminant: D65, A,C,D50, D55, D75, F1, F2 (CWF),F3, F4, F5, F6, F7 (DLF), F8, F9, F10 (TPL5) F11 (TL84),F12 (TL83/U30)
Display Contents: Spectral Value/Graph, Colorimetric Value, Color Difference Value/Graph, PASS/FAIL Result, Color Offset
Measurement Time: About 1.2s(If measure SCI/SCE same time, about 3s)
Repeatability: Spectral reflectance:MAV/SCI,standard deviation within 0.1%(400~700nm:within 0.2%);
Colorimetric value:MAV/SCI,within ΔE*ab 0.04(the average value of whiteboard which was measured 30 times at intervals of 5 seconds after calibration)
Inter Instrument Agreement: MAV/SCI,within ΔE*ab 0.2(Average value of 12 pcs BCRA II series color tiles)
Measurement Mode: Single Measurement,Average Measurement(2~99 times)
Locating Mode: Camera view locating system
Dimension/Net Weight: 184mm×77mm×105mm (L×W×H) /600g
Battery 4 pcs No.5 alkaline batteries (AA alkaline battery); or USB interface as power.
Lamp Life: 5 years, more than 1.6 million measurements
Display Screen: TFT 3.5inch, Capacitive Touch Screen
Interface: USB/RS-232
Data Memory: 1000 Standards, 20000 Sample(A data can include both SCI and SCE)
Operating Tem. Range: 0~40℃,0~85%RH(No condensation),Elevation: Below 2000 m
Storing Tem. Range: -20~50℃,0~85%RH(No condensation)
Standard Accessory: 4 pcs No.5 alkaline batteries,Data line,Operating Instruction,CD-ROM(containing management software)、White and black calibration cavity,Protective cover