Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Column Oven:
Temperature range: Ambient temperature +7℃ ~ 400℃ (in 1℃ increment)
Temperature accuracy: ± 0.02℃
Cooling time: 400℃ to 50℃ in 8-10 min at 25℃ ambient
Programmed temperature setting: 0.1℃~40℃/min (in 1℃ increment)
Program ramps: 7 ramps in total (10 ramps available with control workstation)
Size (L*W*H): 284*280*241mm(inside); 340*345*281mm(outside)
Flame Ionization Detector (FID):
Detection limit: ≤3×10-12g/s (C16)
Best test result: ≤3×10-12g/s (C16)
Drift: ≤6×10-13A/30min
Noise: ≤5×10-14A
Linearity range: ≥106
Thermal Conductivity Detector (TCD):
Sensitivity: ≥5000mV, mL/mg (C16)
Baseline noise: ≤20µV
Drift: ≤60µV/h
Linearity range: ≥104
Flame Photometric Detector (FPD):
Detection limit: ≤8×10-13g/s(P) ; ≤8×10-11g/s(S)
Drift: ≤2×10-11A/30min
Baseline noise: ≤5×10-12A
Nitrogen-Phosphorus Detector (NPD):
Detection limit: ≤5×10-12g/s(N) (Azobenzene); ≤5×10-13g/s(P) (Malathion)