Direct access to our certified experts

Temperature Range
Temperature Range ( @ 0ºC ambient): -63 to 150 ºC
Temperature Range ( @23ºC ambient): -50 to 150º C
Accuracy with (internal reference sensor): ±0.15ºC
Accuracy with external reference sensor ( TC version only) 12 months, relative to external reference sensor: ±0.07ºC
Stability: ±0.02 ºC
Axial Uniformity ( lower 40 mm): ±0.08ºC
Radial Uniformity ( hole to hole): ±0.02 ºC
Calibration Volume
Insert Diameter: 35 mm
Block Depth: 135 mm
Standard insert supplied: 1 insert with 3.5, 4.5,5.5, 6.5, 8.5.10,5 mm pockets
Display
Range: -70 to 150ºC
Type: Led dual line
Resolution: 0.1ºC/0.01C user selectable
Temperature unit: ºC/ºF
Heating time ( including stabilization time): Ambient to 150 C in 30 minutes
Cooling Time ( including stabilization Time): Ambien to –50 C in 60 minutes
Inputs
Switch test: With automatic log of switch open and switch closed
Thermocouple Inputs ( TC version only)
Type: K, N, R,S,T,J,E
Accuracy ( Include CJC error): ± 1 ºC
Display Resolution: 0.1 or 0.01 ºC
RTD’S input and Reference Input ( TC version only)
Type: PT100 IEC 3/4 wires
Accuracy: ± 0.3 ºC
Display Resolution: 0.1 or 0.01 ºC
Digital Interface: RS232 as standard, optional USB converter
Mains: 115 or 230 VAC, 50/60 Hz
Power: 300 VA
Weight: 8 KG
Dimension: 160x340x330 mm