General
Sampling points: 1-10
Display: 3.5¨ TFT color graphical display, back light, window protected
Keyboard: Tactile keypad password protected menu
Industrial Interfaces/alarms/communications/Data logging
Analogue: Up to 40 x 4-20 mA output/input analogue channels (self powered), individually configurable for each gases, with hold last reading function and delay resume function (user selectable)
Alarms: Up to 20 user configurable alarm which trig a relay output (free contact 24VDC/5A) on above/below condition
System Alarm: 1 system alarm
Digital: Modbus RTU as standard, optional ethernet, profinet, profibus dp and web enabled remote access
Data logging: Internal extensive data logging function with user define logging interval- Data export to SD card
Event logging: System monitoring with extensive event logging of faulty conditions to SD card for trouble shooting
Sample
Gas inlet: Stainless steel gas fittings with 1/8” ID thread
Cooler: Integrated Peltier cooler with continuos condensate draining pump
Filter: Teflon particulate filter, Internal Viton hosing
Flow: Monitored and regulated flow 40...60 l/h
Pressure: Sample Inlet Pressure -100 mbar trough 300 mbar, auto regulated
Safety
Norms: Comply to EN650079-15 and RL 94/9/EG
Case Sniffing: Internal cabinet gas sniffer with alarm and power supply cut off (Opntional)
Flow: Stainless steel flow orifice flow restrictor
Ventilation: Ventilated cabinet with monitored fan
Physical
Dimension/Weight: 700 x 600 x 210 mm (H x W x D) wall mounting, 45 KG
IP RATE: IP 54 Aluminium
Notes: * Accuracy of the calibration gas not included. When calibrated with-in its recommended service interval