FLUKE FLUKE-724/APAC/EMEA Temperature Calibrator (RTDs và Thermocouples)
Manufacturer: FLUKE Model: FLUKE-724/APAC/EMEA P/N: 5131125 Origin: China - Contact
- Source/Measurement Voltage DC
+ 30.000 V, 20.000 V, 100.00 mV/0.02%+ 2 counts
+ -10.00 mV to 75.00 mV/0.025 % + 1 count
- Measurement Current DC: 24.000 mA/0.02%+ 2 counts
- Source/Measurement Resistance
+ 0.0 to 400.0 Ω/0.1Ω (4-wire), 0.15Ω (2- and 3-wire)
+ 401 to 1500 Ω/0.5Ω (4-wire), 1Ω (2- and 3-wire)
+ 1500 to 3200 Ω/1Ω (4-wire), 1.5Ω (2- and 3-wire)
- RTDs and Thermocouples: NI-120, PT100, PT200, PT500, PT1000, J, K, T, E, R, S, B, L, U, N, XB, BP
- Loop Power Function: 24V/Max22mA
Batteries: Four AA alkaline batteries.
Size: 96 x 200 x 47 mm
Weight: 650 g
including:
AC70A Test Clips
One pair of stackable test leads
TL75 Test Leads
CE and CSA Markings
Manuals in English, French, German, Spanish, Italian, Portugese, Korean, Chinese and Japanese
Users manuals on CD-ROM
NIST- traceable calibration report and data
Includes:
AC70A Test Clips
One pair of stackable test leads
TL75 Test Leads
CE and CSA Markings
Manuals in English, French, German, Spanish, Italian, Portugese, Korean, Chinese and Japanese
Users manuals on CD-ROM
NIST- traceable calibration report and data
Datasheet
Specifications: Fluke 724 Temperature Calibrator
| Measurement Accuracy | ||
| Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
| 20.000 V | 0.02% + 2 counts (lower display) | |
| 100.00 mV | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA | 0.02% + 2 counts |
| Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
| 1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
| Source Accuracy | ||
| Voltage DC | 100.00 mV | 0.02% + 2 counts |
| 10.000 V | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) | |
| Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
| 1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
| RTDs and Thermocouples | ||
| Measure accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.8°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Source accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| Accuracy stated for 4-wire measurement. | ||
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.4°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Specifications | ||
| Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
| Loop power function | Voltage | 24 V |
| Accuracy | 10% | |
| Maximum current | 22 mA, short circuit protected | |
| Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Steps | 25% of range, 100% of range | |
| Environmental Specifications | ||
| Operating temperature | -10°C to 55°C | |
| Storage temperature | -20°C to 71°C | |
| Humidity (without condensation) | 90% | 10°C to 30°C |
| 75% | 30°C to 40°C | |
| 45% | 40°C to 50°C | |
| 35% | 50°C to 55°C | |
| Safety Specifications | ||
| Safety rating | CSA C22.2 No. 1010.1:1992 | |
| EMC | EN50082-1:1992 and EN55022:1994 Class B | |
| Mechanical and General Specifications | ||
| Size | 96 x 200 x 47 mm | |
| Weight | 650 g | |
| Batteries | Four AA alkaline batteries | |
| Battery life | 25 hours typical | |
| Shock and vibration | Random, 2 G, 5-500 Hz | |
- Quality Engagement
- Easy change and return
- Delivery Avaliable