Dust technical parameter

Sampling flowrat: Range: (0~110)L/min, Resolution: 0.1L/min, Error: ±2.5%

Flowrate control: ±2.0%(voltage change ±20%,resistance range:3kpa—6kpa)

Dynamic pressure: Range: (0~2000)Pa, Resolution: 1Pa, Error: ±1.0%FS

Static pressure: Range: (-30~30)kPa, Resolution: 0.01kPa, Error: ±1.0%FS

Total pressure: Range: (-30~30)kPa, Resolution: 0.01kPa, Error: ±2.0%FS

Flowrate pre-meter pressure: Range: (-60~0)kPa, Resolution: 0.01kPa, Error: ±1.0%FS

Flowrate pre-meter temperature: Range: (-55~125)℃, Resolution: 0.1℃, Error: ±2.5℃

Velocity range: Range: (1~45)m/s,  Resolution: 0.1m/s,  Error: ±4.0%

Atmospheric pressure: Range: (60~130)kPa, Resolution: 0.1kPa,  Error: ±0.5kPa

Auto tracking Precision: Range: -, Resolution: -, Error: ±3%

Max sampling volume: Range: 99999.9L, Resolution: 0.1L, Error: ±2.5%

Isokinetic trackingresponse time: ≤10s

Load capacity of pump: ≥50L/min(when resistance is 30 PA)

Size: (L275×W170×H265)mm

Weight: About 6.8kg(battery include)

Noise: <65dB(A)

Power consumption: <300W


Flue gas technical index

Sampling flow: Range: 1.0L/min, Resolution: 0.1L/min, Error: ±5%

O2(optional): Range: (0~30)%, Resolution: 0.1%, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

SO2(optional):  Range: (0~5700)mg/m3  Can extend to 14000mg/m³,  Resolution: 1mg/m3, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

NO(optional): Range: (0~1300)mg/m3  Can extend to 6700mg/m³,  Resolution: 1mg/m3, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

NO2(optional):  Range: (0~200)mg/m3  Can extend to 2000mg/m³, Resolution: 1mg/m3, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

CO(optional): Range: (0~5000)mg/m3  Can extend to 25000mg/m³, Resolution: 1mg/m3, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

H2S(optional): Range: (0~300)mg/m3  Can extend to 1500mg/m³, Resolution: 1mg/m3, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

CO2(optional):  Range: (0~20)%, Resolution: 0.01%, Error: Indication error:±5%Repeatability:≤1.5%Response time:≤90sStability:change within 1hExpected life:2 years in air (beside CO2)

Details

Product Detail

It adopts isokinetic sample and membrane filter (cartridge) weighing method to measure dust concentration while using electrochemistry or optical principle sensor  to analyze O2, SO2, CO, NO, NO2, H2S, CO2. And also the velocity of flue gas, flue gas temperature, flue gas humidity,  flue pressure and exhaust air rate etc.

Standards

EN13284-1

US EPA M5

US EPA M17

ISO 9096