Measurement Type: Absolute
Sensor Temperature: Unheated or controlled to 45°C, 80°C or 100°C
Full Scale Pressure Ranges: 0.1, 0.25,1, 2, 10, 20, 100, 200, 500, and 1000 Torr
Accuracy: Unheated: 0.25% Reading
45°C Models: 0.12% Reading for 1-1000 Torr, 0.15% Reading for<1 Torr
80°C and 100°C Models: 0.25% Reading for 1-1000 Torr, 0.50% Reading for<1 Torr
Communication Protocol: DeviceNet™
Analog Output; 0-10 VDC into >10KΩ load
Analog Connector: 9-pin female D-subminiature
Digital Connector: 5-pin sealed micro-style male bayonet with anti-rotation device
Span Temperature Coefficient: Unheated Models: 0.04% Reading/°C
Heated Models: 0.02% Reading/°C
Volume: 6.3 cc
Overpressure Limit: 45 psia (310 kPa)
Exposed Materials: Inconel and Incoloy nickel alloys
Power Requirements: Unheated Models: 11-25 VDC @≤5 watts
45°C Models: 18-25 VDC @≤15 watts
80°C & 100°C Models: 18-25 VDC @≤25 watts
Level of Filtering: Adjustable via user software
Data Rate Switch: 4 positions: 125, 250, 500, PGM
Network Size: 64 nodes maximum
Compliance: CE
For specific range please check the attachment below!