Measurement Technique: Mass Spectrometry
Mass Range: 1-100 amu
Detector: Faraday
Ion Source: Open
Filaments: Tungsten
Software: Built-in web application
Filter length: 4 in. (100 mm)
Maximum Analyzer Operating Pressure: 1e-4 Torr (1.3e-4 mbar)
Ion Source Sensitivity: 2e-4 A/mbar
Mass Stability: ±0.1 amu over 8 hours at stable ambient
Resolution: Better than 10% valley for peaks of equal height across mass range
Bake Out Temperature: 250°C with electronics removed
Maximum Operating Conditions: 10 - 40°C, 80% RH (non condensing)
Power: 24 VDC, 2A external universal voltage supply (included)
Settling Time: <20 ms
Control Method: CE operating system processor with web-server interface
Compliance: CE