Number of channels: 4
Output Type: Standard digital output
Probe Configuration: Push/Pull configuration
Measurement Range : 0 to 12.5mm
Noise: 0.000001√Hz FSR
Repeatability : 0.000085% FSR (at a fixed point, 1 Hz bandwith)2
Minimum System Resolution : 0.100 nm
Long Term Stability/Drift : 20ppm /month or better at (±1° C)
Linearity Accuracy : ±0.01% FSR
Frequency Response : 5kHz
Output Data Rate : 100 min. to 20,000 max. (samples per second)
Temperature Stability 1: 00 ppm digital (over 0 to 40°C)
Butterworth Filter : 50, 100, 500, 1kHz, or 5kHz
Range Extension : 1x and 2x Default. Up to 10X max. optional (see probe charts for max probe range extension permissable)
ADC Bit Count : 24-bits
Exponential Filter : No Filtering, 0.1 , 1 or 10 Hz
Basic Interface : Command-Response, ASCII commands
Digital Output : Micro USB or RJ-45 Ethernet 10/100/1000
Analog Output Span : 0-5V (14 bit resolution), 0-10V(15 bit resolution), -10V to +10V (16 bit resolution), -5V to +5V (15 bit resolution)
Analog Output Impedance : 50 Ω, 5kHz, 5 pole Butterworth Low Pass Filter Limited
Encoder Input : 0-24VDC max, Threshold ~1.2 V, 32 bit, Z input/ reset input
Included Software : MTI Basic Software, LabVIEW, .NET, and DLL Drivers
Operating Temperature : 0 – 40°C, 95% non-condensing (designed); 20°C, 100kPa, 50%RH (nominal)
Operating Environment : IP40 (particles to 1mm/ no water protection)
Power Requirements : 24VDC±1V 50mV ripple, switching speed >60kHz. <8W estimated
Target Ground Return : Integrated with Power Connector
Input Protection : Reverse Polarity (Over Volt to 35VDC)
ESD Protection : ±4kV Contact and ±8kV Air
Case Dimensions : 2” (53mm) H × 4” (103mm) W × 4.7” (120mm) D
Case Mount : DIN Mount Kit
Probe Connectors : SMA Female