Metrological parameters:
E1 Calibration Range : 1 – 20 kg
E2 Calibration Range : 1 – 20 kg
F1 Calibration Range : 1 – 20 kg
F2 Calibration Range : 1 – 20 kg
Maximum capacity [Max] : 26.1 kg
Readability [d] : 1 mg
Standard repeatability [5% Max] : 2 mg
Standard repeatability [Max] : 3 mg
Permissible repeatability : 6 mg
Electric compensation range : 100 g – +26.1 kg
Stabilization time : 30 s
Adjustment: external
Physical parameters:
Display : 10″ graphic colour touchscreen
Weighing pan dimensions: ø220 mm self centering
Object diameters : 25 – 145 mm
Electrical parameters:
Power supply : 100 – 240 V AC 50/60 Hz
Environmental conditions:
Operating temperature : +15 – +30 °C
Storage temperature : -20 – +50 °C
Operating temperature change rate : ±0.5 °C / 12 h (±0.3 °C / 4 h)
Relative humidity : 40% – 60%
Relative humidity change rate : ±5% / 12 h (3% / 4 h)