For full functionality of this site it is necessary to enable JavaScript.
EMIN.COM.MM
0

Scale Calibration Procedure

04/26/2021 13:44:25

**1\. Range of application** This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not gre

1. Range of application

This technical document specifies a calibration procedure for scale (non-automatic scales class I) with a division value not greater than 1 mg. Technical scale is (non-automatic scale class II) according to OIML R76.

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

- Analytical Balance:
Sets of weights E2, Class E2, Measuring range: 1 mg ÷ 500g  
Sets of weights F1, Class F1, Measuring range: 1g ÷ 500g
Kern ADB 200-4 Analytical balance (200g/0.1mg)


- Technical Scale:
Sets of weights F1, Class F1, Measuring range 1g ÷ 500g
Standard Weight F1, Class F1, Measurement weight: 1 kg, 2 kg, 5 kg, 10 kg.


- Bench Scale: Standard Weight M11, Class M1, Measuring range 0 ÷ 150kg



3.2 Auxiliary Instrument(s)
Hygro - Thermometer:

- Range of Measurement: (0 ÷ 50) °C
- Accuracy: ± 0,5 °C
- Range of Measurement: (10 ÷ 95) %RH
- Accuracy: ± 3 %RH

4. Conditions for Calibration
The calibration site must be bright, away from heat sources, wind sources, and not vibrated.
Temperature, humidity in the calibration site ensures:
Ambient temperature: (22 ± 3) ° C
Air humidity: (40 ÷ 80) %RH.
The scale should be calibrated at the site of use.

5. Prepare for Calibration
Before calibration, the following preparations must be done:
- The scale must be clean.
- Place the scale to be calibrated firmly on a surface and adjust the scale.
- Turn on the scale for at least 30 minutes or as required by the manufacturer.
- The scale to be calibrated should be thermally equilibrium with the environment in which it is calibrated. The minimum time required for this process is specified by the manufacturer.

6. Begin the calibration
6.1 Outside check
6.2 Technical Check
6.3 Measurement check
6.3.1 General requirements
6.3.2 Begin the calibration
- Test repeatability
- Check the deviation angle
- Check the correctness

7. Measurement Uncertainty Evaluation
8. General handling
- The calibrated scale is stamped, certified in paper.
- The calibration result includes standard value, indicator value, main number, together with the measurement uncertainty expanded for each weight level, coverage factor k.
- The recommended calibration cycle of scale is 12 months.

9. About EMIN's Calibration - Verification - Repair center
EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)
If you want your instruments calibrated, please contact:
Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 
Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City
Email: services@emin.vn - Hotline: 0968.060.220
Further information about scale calibration service can be found here.


Related News

SMEMA standard for PCB conveyors
11/16/2026 14:00:39

SMEMA was developed to address this specific point; consequently, conveyors, loaders, and unloaders all include an SMEMA port in their technical specifications

Homes on busy streets: How to measure fine particulate matter and exhaust fumes entering your living room
11/16/2026 09:42:31

Living in a house facing a main road is particularly frustrating: you might wipe the tables until they’re spotless in the morning, only to find them covered in a layer of dust by the afternoon. Even with the doors tightly shut and the curtains drawn, the house still feels stuffy by evening—permeated by the faint, lingering presence of vehicle exhaust. While regular dust can be swept away, fine particulate matter the kind you can’t even see silently seeps in through the gaps around your doors and windows. So, before you rush to spend money on an air purifier or weather-stripping, take a moment to check the current air quality inside your home and identify exactly where the pollutants are entering

Can Laser Welding Machines Weld Aluminum Effectively ?
11/16/2026 08:54:12

Laser welding is presented as an aluminum-joining method that concentrates heat at the weld point, helping limit distortion. Suitable settings depend on material thickness, filler wire use, and trial weld results.

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts