Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Rated Output (0℃-40℃):
Channel/Range: Channel1; Channel2/ Channel3/ Channel4
- Voltage: 0-30V/ 0-10V/ 0-5V
- Current 0-3A/0-3A/0-1A
- Power: 90W/30W/5W
Load regulation ± (% Output + Bias):
- Voltage: ≤0.01% + 3 mV
- Current: ≤0.01% + 3 mA
Power regulation ± (% Output + Bias):
- Voltage: ≤0.01% + 3 mV
- Current: ≤0.01% + 3 mA
Programming resolution:
- Voltage: 1mV
- Current: 0.1mA
Read-back value resolution:
- Voltage: 1mV
- Current: 0.1mA
Year accuracy (25℃± 5℃) ± (% Reading + Bias):
- Programming:
Voltage ≤0.1% + 20 mV
Current ≤0.2%+5mA
- Readback
Voltage: ≤0.1% + 20 mV
Current: ≤0.2%+5mA
Ripple and Noise (20Hz-20MHz):
- Normal mode voltage: ≤1mVrms or 3mVp-p
- Normal mode current: ≤3mArms
Series and parallel set value accuracy:
- Voltage: ≤0.02% + 10 mV
- Current: ≤0.2% + 20mA
Timer: 0.1 ~ 99999.9 seconds
Memory: 40 groups of settings files / channels